Glove Box Regeneration Measures

From HealthTimes
Jump to: navigation, search

1) Regeneration fuel: a mixture of hydrogen and inert gasoline (hydrogen accounts for five% -10%)

two) Amount: 40L gas cylinder 10Mpa

three) Regenerative gas strain: .05Mpa (the gas consumption for each 50 percent hour in the point out of .05Mpa is that the cylinder force drops by 1.5Mpa. If it exceeds one.5Mpa, the stress of the stress regulator’s program will be minimized, if it is less than 1.5Mpa Increase the stress of the stress regulator plan)

4) System: 3 hrs heating-3 hrs heating and ventilation-three several hours vacuuming-1 hour pumping-ten several hours cooling

five) Techniques:

① Connect the regeneration gasoline (test the pipeline for leaks right after connection: following connecting, open the main valve of the force minimizing valve to the maximum, open up the auxiliary meter to .1Mpa, change off the principal valve, wait around fifteen minutes, if the tension of the auxiliary meter stays at .1 Mpa remains unchanged, indicating no air leakage

>②Set the tension of the auxiliary gauge of the stress cutting down valve to .1Mpa (the strain throughout air flow is about .06-.08MPa

>③ Shut the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Start off regenerat

br>⑥ The new software ends at 20h (thoroughly clean just before cleansing, water and innovative technology glovebox oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.